Each type of membrane displays a different affinity for various molecules. For globular proteins, the relative binding affinity is CE < RC < PVDF.
Standard RC membrane (Spectra/Por® 1-7) and Biotech RC is constructed of flexible regenerated cellulose polymers and can be sealed using any of the dialysis tubing closures. Biotech CE and PVDF are constructed of a more rigid polymer requiring gentler Universal closures. Since these work well for all dialysis tubing; when in doubt, use Universal Closures. Standard Closures should ONLY be used with Standard RC tubing.
It is recommended to use a closure with a sealing width of 4-10 mm longer than the flat width of the dialysis tubing. The smallest Universal Closure has a sealing width of 50 mm. This will seal all flat widths of our Biotech Grade tubing.
The dry packaged dialysis membranes have a shelf-life of 5 years. The wet packaged (0.05% sodium azide solution) membranes have a shelf-life of 3 years. The irradiated membranes have a shelf life of 1.5 years.
If wetted membrane dries out, the pore size is adversely affected and the membrane becomes brittle and will likely leak. The membrane should be discarded.
If the membrane freezes, the ice crystals may rupture the membrane and also cause leaking. It is recommended not to use the membrane. However, you can try slowly increasing the temperature until the storage solution completely melts. The possibility remains that the membrane is not longer integral.
No. Repligen manufactures dialysis membranes intended for laboratory use.
The roller marks or fold lines along the membranes will not affect the diffusion properties as long as the membrane is integral.
While the polymers in the cellulose ester (CE) membrane cross-link to form a more rigid molecular lattice, the regenerated cellulose (RC) polymers form a more flexible lattice structure. The opaqueness comes from the pores in a more rigid frame. The larger the pore size, the more opaque the membrane.
The CE and RC dialysis membranes can only be mechanically sealed. However, the PVDF dialysis membranes can be mechanically sealed or heat sealed and is often used in this manner for the purposes of sample "encapsulation".
We do not recommend re-using dialysis membranes since they can be contaminated through handling and dialysis conditions (pH, temperature, chemical exposure, etc.) can alter the membrane integrity and/or cause leaking, especially when removing and reapplying closures. Dialysis membranes are designed for single use.
Since dialysis membrane consists of a spongy matrix, it is more appropriate and practical to measure the "pore size" indirectly by rating its retention performance characterized by its Molecular Weight Cut Off (MWCO). The MWCO is defined by the molecular weight solute that is 90% retained by the membrane during a 17 hour period. For this reason, you should select a MWCO that is just smaller than the size of the solutes you want to retain.
Both Spectra/Por® 2 and 4 have a MWCO of 12-14 kD. While Spectra/Por® 4 is more suited for general dialysis, Spectra/Por® 2 offers special higher and lower FW's and/or higher permeability; i.e. water permeability of Spectra/Por® 2 is superior to that of Spectra/Por® 4.
The surface area to volume ratio is a function of the tubing flat width. If you have a two equal length pieces of tubing with two different flat widths, the smaller flat width piece possesses a higher surface area to volume ratio and dialyzes quicker while the larger flat width piece possesses a lower surface area to volume ratio and dialyzes slower. The smaller flat width has a shorter distance for diffusion and less solute competition through the membrane pores. Larger flat widths have a longer distance to the membrane and more solute competition through the pores. In general, the greater the surface-area-to-volume ratio, the quicker the dialysis.
Most dialysis is done with no osmotic pressure across the membrane. The dialysis process is driven by the concentration gradient from the inside and outside of the dialysis tubing. If there is a large difference in osmotic pressure, water will move across the membrane. If too much water migrates across the membrane, the dialysis tubing can potentially burst or collapse, depending upon the direction of water movement.
Dialysis membranes are not designed for pressure filtration. The maximum recommended pressure is 1.5 psi without affecting the MWCO.
Biomolecules must be maintained under strict pH control to stabilize their molecular properties. The typical pH range for dialysis buffers is 6 to 8. The following are some of the common solutions/buffers found in biochemical solutions:
While the size of dissolved molecules is defined by molecular weight (MW) units in Daltons, particle and cell size is defined by metric diameter because the MW units become impractical and do not account for shape in the microscopic range. Since microns are a measure of a 2-dimensional distance and Daltons are a measure of 3-dimensional size based on atomic weight units, there is no direct conversion from one to the other. For this reason, many common biological materials were characterized for dialysis, ultrafiltration and microfiltration purposes and plotted on a conversion chart to correlate the approximate scales as a reference for estimating conversions. Refer to Repligens' Pore Size Chart on our website for converting between Daltons and metric units.
We recommend performing dialysis against a low salt concentration buffer if possible. It is also not advisable to dialyze against pure DI water as this will cause the osmotic pressure to draw water into the tubing, "balloon" and potentially rupture the membrane. A serial dialysis using buffers with decreasing concentration of solutes (salt) will prevent the osmotic pressure from swelling the membrane. Try to reduce the order of magnitude of solute concentration by a factor of 10 to 1 at each buffer exchange; i.e. dialyze a sample with 5 M NaCl against a buffer with 500 mM NaCl.
We strongly recommend against tying knots in the tubing. Tied knots do not provide an effective seal against leakage. Only dialysis closures will provide the adequate seal necessary for safe dialysis of your sample.
Along with each tubing flat width, Repligen lists the correlating volume/length ratio that can be used to calculate how much length is required to contain your sample volume. For example if the FW is 16 mm, the volume/length ratio is 0.79 ml/cm. To contain a sample of 5 ml, you will need a length of approximately 6.5 cm. However, you also need to add about 10 to 20% (but at least 1 cm) more length to account for head space (air) to keep your sample buoyant. Lastly, you will need to add enough length, about 2 cm at each end, to allow for applying two closures. The total tubing length would be at least 11.5 cm. The simple equation to calculate total required tubing length is as follows:
Total length = (sample volume) / (vol/length) + (additional 10-20%) + 4 cm