Variable Pathlength Spectroscopy for Accurate Protein Concentration Measurement Using A280

Protein Concentration Measurement using A280 Overview

Protein concentration determination is integral to in-process control throughout biomanufacturing to monitor and optimize the production of proteins, antibodies, and other biologics. Absorbance at 280 nm, or A280, is the industry-standard method for determining protein concentration. Whether for process development, manufacturing, or quality control, accurate A280 measurements are essential for various bioprocess steps such as purification, ultrafiltration and diafiltration (UF/DF), formulation, and release testing.

A280 is an application of the Beer-Lambert law, A=εlc. If the protein’s extinction coefficient ε is known and the pathlength l is kept constant, then absorbance A is linearly proportional to concentration c.

Current Challenges

Transitioning to Continuous Processing

As bioprocessing practices transition from batch to continuous processing to enhance efficiency and productivity, there's a critical need for advanced methods for quantifying protein concentrations. Existing methods face two main challenges: measuring high protein concentrations and measuring concentration in real time. 

Challenges with Traditional Spectrophotometers

New protein therapeutics, whether investigational or commercial, frequently demand high concentrations to enable effective subcutaneous administration. Traditional spectrophotometers, which use a fixed pathlength for A280 measurement, do not have a sufficient dynamic range to measure the concentration accurately at all stages of the process. Dilution techniques can bring the sample within the detectable range, but this increases the chances of variability or error due to differences in operator skill levels, preparation procedures, sample viscosity, and the need for meticulous scale calibration. 

Limitations of In-line UV-Vis Sensors

In-line UV-Vis sensors provide an integrated approach for real-time monitoring of protein concentration. However, these sensors face their own challenges, notably oversaturation and limited dynamic range, especially when dealing with high-concentration biologics. The high absorbance levels of concentrated samples can surpass the sensor's detection capabilities, resulting in signal saturation and inaccurate measurements. Furthermore, the inherent limitations in dynamic range may compromise the sensor's ability to accurately capture variations in concentration across a wide spectrum, posing challenges for precise monitoring. Given the in-line nature of these sensors, dilution strategies are not viable, exacerbating the need for robust solutions to address these challenges and ensure accurate in-process monitoring of protein concentration.

The Need for Innovative Analytical Solutions

As the biopharmaceutical industry progresses towards more potent drug formulations, traditional fixed-pathlength methods struggle to meet the analytical requirements. The emergence of stronger, smaller, and more targeted batches of drugs, supporting a personalized medicine approach, underscores the need for innovative analytical solutions.

Advantages of Variable Pathlength Spectroscopy for Protein Concentration Measurement (A280)

Variable Pathlength Spectroscopy operates within the framework of the Beer-Lambert Law, A=εlc. However, unlike traditional approaches where the pathlength l remains fixed, Variable Pathlength Spectroscopy strategically varies the pathlength, thereby unlocking the constraint on l. This variability in pathlength enables a wider dynamic range in concentration measurement, eliminating the need for dilutions and allowing the concentration c to remain fixed during measurement. 

In this method, light enters the system through a fiber optic cable at the top of the instrument, passes through a solid Fibrette into the sample, and exits through the bottom. The pathlength is the distance between the bottom of the Fibrette and bottom of the sample vessel, which is adjusted by a stepper motor for precise control over the pathlength variation.

The CTech™ SoloVPE® and FlowVPX® Systems harness the power of Variable Pathlength Technology (VPT) to deliver rapid and precise A280 measurements for protein concentration. By utilizing VPT, our systems offer numerous advantages over traditional UV-Vis spectrophotometers. They conduct multiple absorbance measurements per sample, enabling slope-based concentration results that yield superior accuracy compared to conventional methods. This approach eliminates errors associated with additional dilutions and calculations and streamlines the measurement process. Pathlength adjustment increases the dynamic range of the system, granting the ability to measure concentrations from 0.5 to 300 mg/ml, depending on the molecule of interest, with impressive accuracy within 1%. VPT systems also include built-in quality checks to ensure reliable and consistent results across a wide concentration range.

VPT systems are available in two options for bioprocessing analytics:

CTech™ SoloVPE® System

CTech™ SoloVPE® System provides at-line concentration values in less than 2 minutes, with no need to send samples off to QC labs for process-disrupting concentration checks.

  • VPT allows the direct measurement of proteins, including monoclonal antibodies (mAbs), which absorb light at 280 nm due to aromatic amino acids (primarily L-tyrosine and L-tryptophan).
  • The system takes 5 to 10 absorbance data points, generating a slope regression and ensuring R2 ≥ 0.999.
  • System-to-system consistency and streamlined and fast implementation and validation.

CTech™ FlowVPX® System

CTech™ FlowVPX® System delivers in-line, real-time concentration measurement with continuous reading throughout the process.

  • Provides accurate measurement during the downstream bioprocess from 0.1 mg/ml to more than 250 mg/ml for an antibody with ε = 1.5 ml/(mg·cm). 
  • No need to remove samples for testing or disrupt production flow. 
  • In-line measurements using the FlowVPX system can reduce hold times in production and allow for better process automation. 
  • Utilized in various stages of the downstream bioprocessing workflow.

ViPER® ANLYTX Software

CTech™ ViPER® ANLYTX Software is a next-generation software platform for all your protein concentration measurement (A280) needs. ViPER software features an app-based architecture, permitting easy interpretation of sample data. The software automatically generates key insights from calculations of absorbance over pathlength and displays slope, R2, wavelength, concentration, extinction coefficient (if known), and other user-specified results.

With ViPER software, information enables action.

Expert Webinar Series

The "Go with the Flow" program delves into the transformative power of variable pathlength spectroscopy in enhancing Process Analytical Technology (PAT) for biopharmaceutical development and manufacturing. Join industry as they share insights on implementing in-line PAT solutions, such as CTech FlowVPE/FlowVPX technology, to drive process innovation and efficiency. Discover how these advanced technologies revolutionize protein concentration monitoring, enabling real-time control and automation in biologics production.

Watch now

Your success is our success

Application Services

CTech is committed to our customers’ success. We design service offerings to supplement our standard support options and provide you with increased access to our highly knowledgeable and experienced professionals. We offer implementation, development, and educational guidance to maximize the benefits of your variable pathlength solutions and the Slope Spectroscopy® technique.

Application Services

  • 21 CFR Part 11 Guidance
  • Annex 11 Guidance
  • Validation Support
  • Qualification Service
  • Method Transfer Service

Development

  • Method Design & Development
  • Comparability
  • On-Site Support
  • Custom Standards
  • Process Auditing

Education

  • SOP Document Review
  • Advanced Slope Spectroscopy Training
  • Data Analysis
  • Maintenance & Troubleshooting

Applications

Drug-to-antibody Ratio (DAR)

The SoloVPE System provides accurate, precise, and robust concentration measurements of Antibody-Drug Conjugates (ADCs) and Drug-to-Antibody Ratio (DAR).

Read More

Gold Particle Testing

SoloVPE systems can be used for Gold particle testing, a highly effective method to validate filters and process integrity in biopharmaceutical manufacturing

Read More

Multicomponent Analysis

SoloVPE System’s variable pathlength technology allows multiple hemoglobin components to be measured simultaneously without the need for baseline correction, scatter correction, or dilution

Read More

Polysorbate Analysis

PATsmart™ SoloVPE System enables the PS80 and PS20 stock solution to be measured without dilution and highly reproducible results through variable pathlength technology.

Read More